Control of sensitivity in metal oxide electrolyte gated field-effect transistor-based glucose sensor by electronegativity modulation | Scientific Reports
Scientific Reports volume 14, Article number: 27251 (2024) Cite this article
Metrics details
In this study, the sensitivity of electrolyte-gated field-effect transistor-based glucose sensors using oxide semiconductor materials was controlled via electronegativity modulation. By controlling the enzymatic reaction between glucose and glucose oxidase, which is affected by the surface potential, the sensitivity of the glucose sensor can be effectively adjusted. To evaluate the sensitivity characteristics of the glucose sensor according to electronegativity control, devices were fabricated based on InO through Ga and Zn doping. The results confirmed that the specific sensitivity range could be adjusted by increasing the electronegativity. In addition, density functional theory calculations, confirmed that the attachment energy of the surface-functionalized material and the enzyme binding energy in the surface-functionalized thin film can be modulated depending on the electronegativity difference. The dissociation constant was controlled in both directions by doping with metal cations with larger(Ga, 1.81) or smaller(Zn, 1.65) electronegativities in InO(In, 1.78). We expect that this study will provide a simple method for the gradual and bidirectional control of the glucose sensitivity region.
Diabetes is a rapidly growing public health problem due to the exponential increase in obesity worldwide1,2,3. Glucose plays a key role in human metabolic processes, and abnormal glucose levels in the blood or body fluids are directly related to diabetes. Therefore, it is important for diabetics to constantly monitor their blood sugar levels (2–20 mM range)4,5. This monitoring involves performing a puncture test to extract blood, which can be painful6. The pain generated by this invasive method causes the patients to avoid self-control7. Therefore, much attention has been devoted to the noninvasive monitoring of continuous glucose levels. To replace this invasive method, which uses blood glucose, high sensitivity at low glucose concentrations is required, as the glucose concentration in body fluids such as tears (0.05–5 mM), saliva (0.008–1.77 mM), sweat (0.01–1.11 mM), and urine (0–0.8 mM) is much lower than in blood8,9,10,11. In addition, for long-term healthcare monitoring, selectivity, low cost, and high sensitivity tests should be considered12,13.
Metal-oxide semiconductor field-effect transistors have recently received much attention as biosensors owing to their diverse merits, such as easy fabrication methods with solution processes or sputtering systems, good stability and chemical resistance, and easy surface functionalization. Therefore, metal-oxide semiconductor thin films are superior sensor materials for biochemical detection14,15. In particular, electrolyte-gated field-effect transistor (EGFETs)-based sensors using metal-oxide semiconductors such as InO16, ZnO17,18, Ta2O5-ZnO19, and InGaZnO-FETs1,20,21 have been intensively studied because of their advantages of stability in aqueous environments, operation at low driving voltages, and ability to convert and amplify biological signals into electronics22. In addition, research on glucose sensors using surface functionalization and/or enzymatic immobilization of metal-oxide thin films is actively being conducted. The advantages of non-enzymatic sensors include lower device fabrication costs and excellent stability. However, they suffer from low selectivity and interference from various compounds23,24. In contrast, when using enzymes, it is possible to operate at low voltages (< 0.6 V) and achieve high selectivity and specificity25. Glucose oxidase (GOx) not only exhibits high selectivity for glucose targets but also withstands a wide range of pH and temperature conditions compared to other enzymes, making it the most common enzyme and a key component in numerous glucose biosensors26. According to several reports, enzymatic attachment to substances can be achieved using simple, convenient, and cost-effective methods23,27. Surface termination and enzyme immobilization according to surface functionalization vary depending on the oxide material and composition, and the surface mechanism changes accordingly. Previous studies16,17,18,19,20,21 on glucose sensors using enzyme immobilization on the surface of oxide materials have confirmed the difference in glucose sensitivity characteristics depending on the selected oxide material, as well as the biosensor applicability of various oxide materials. Therefore, it is necessary to study the effect of the surface mechanism on surface termination and enzyme immobilization, as well as the correlation between its glucose sensor properties.
Therefore, we focused on the change in the surface state according to the electronegativity of the surface mechanism and investigated how the surface immobilization phenomenon according to the electronic structure of the metal oxide thin film affects the characteristics of the glucose sensor. To confirm the change in the surface mechanism depending on the surface conditions of the metal-oxide thin film, thin films and devices were fabricated by modulating the composition of the metal-oxide thin films by doping InO with Zn and Ga. Using these devices, the mechanism of the characteristic change of the glucose sensor was discovered, and experiments and theoretical calculations using density functional theory (DFT) confirmed that this mechanism is related to electronegativity.
(a) Scheme of surface functionalization (b) Topologies of the InGaO, InO, and InZnO thin films before and after surface functionalization. (c) C 1s, N 1s, and O 1s XPS spectra of InGaO, InO, and InZnO thin films before and after surface functionalization. (d) XPS spectra showing the deconvolution of C 1s and N 1s of the surface-functionalized InGaO, InO, and InZnO thin films.
Figure 1a shows the process flow of surface functionalization and enzyme immobilization on metal-oxide thin films using a solution process. Before the solution process, UV lamps (185 nm and 254 nm) were used on the thin film’s surface for hydroxylation. For silanization, the samples were immersed in 4 wt% (3-aminopropyl)triethoxysilane (APTES) diluted with toluene. The samples were immediately rinsed with toluene to remove weakly bonded APTES molecules before the next step. After then 1.5 wt% glutaraldehyde (GDx) was diluted with 1x PBS to cross-link the amine functional group between the surface of the silanized metal oxide thin films and the glucose oxidase (GOx) molecules. Finally, the GOx solution was added dropwise to the metal-oxide thin films over 24 h. A more detailed procedure is provided in the Experimental Section. To confirm the surface states before and after the surface functionalization, the morphologies of the InGaO, InO, and InZnO thin films were measured using scanning Atomic Force Microscopy (AFM) at a scan rate of 0.35 Hz. The three-dimensional (3D) topology images are shown in Fig. 1b. Each image measures 0.5 μm × 0.5 μm. After surface functionalization, the surface roughness Rq of the samples increased from 0.311, 0.322, and 0.326 to 1.70, 1.287, and 1.121 nm, respectively.
The structural properties of the InGaO, InO, and InZnO thin films were determined using high-angle annular dark field (HAADF) scanning transmission electron microscopy (STEM) images with energy dispersive spectroscopy (EDS) compositional mapping, and X-ray photoelectron spectroscopy (XPS) can be found in Fig. S1 and Table S1. The respective thicknesses of the InGaO, InO, and InZnO thin films were approximately 10.08, 13.77, and 9.84 nm, respectively, which are similar to the intended thickness (10 nm). In addition, STEM-EDS confirmed that approximately 10 nm of Si was present on the surfaces of the InGaO, InO, and InZnO thin films. For analysis, Pt was thinly deposited onto the thin-film surfaces. However, during this process, Pt penetrated the organic layer, affecting the appearance of the enzyme attached to the surface in the HAADF image and making it appear thinner. To confirm the presence or absence of elements and changes in the chemical binding states of the metal-oxide thin film surfaces as a result of surface functionalization and enzyme immobilization, as well as the amount of doped Zn and Ga, we used XPS analysis to measure the C, N, O, Si, In, Zn, and Ga spectra. Figure 1c shows the C 1s, N 1s, and O 1s XPS spectra of the InGaO, InO, and InZnO thin films before and after surface functionalization. The lowest binding energy OI peak was located at 529.5 eV and can be attributed to the O2− in the metal oxides, that is, In-O, Zn-O, and Ga-O. The OII peak (531 eV) is related to the O− in metal oxides. The highest binding energy OIII peak (532.5 eV) corresponds to the Si-O bonding states and/or electrolyte oxidation (APTES in Si, O, and O in metal oxides)20,27,28. Si and O were detected on all thin-film surfaces after surface functionalization, indicating that APTES adhered well to the thin-film surface, as shown in Fig. S2 and Table S2. In addition, C and N were detected in all thin films, indicating that they were well formed up to the last step of surface functionalization (Fig. 1d). Generally, all elemental spectra were calibrated using the standard binding energy of carbon, which is equal to 284.5 eV. The C 1s spectra were normalized and curve-fitted with five different Gaussian peaks to obtain the detailed chemical carbon bonding states, which were found to have low binding energies for C-C (284.5 eV), C-N and C = N (285.6 eV), C-O (286.5 eV), and C = O (287.7 eV), and high COO binding energies (288.5 eV)29. The C 1s peak was assigned to the C-N and C = N bonds (285.6 eV), confirming the combination of GDx with GOx on each surface of the metal oxide thin film. When Ga and Zn were mixed into the InO material, the relative chemical bonding states of the C-N peaks increased. The N 1s spectra were normalized and curve-fitted using three different Gaussian peaks of the detailed chemical bonding states of nitrogen, which included peaks corresponding to low binding energy (C = N; 398.6 ± 0.1 eV), medium energy (C-N and (Si-)2N-O; 399.6 eV), and high energy (C-N+; 400.7 ± 0.1 eV)7,30,31. Detailed information on the fitting results is provided in Table S2. The ratios of neutral amine nitrogen (C-N and (Si-)2N-O) to positively charged protonated amine nitrogen (C-N+) decreased to 1:0.18, 1:0.16, and 1:0.15 as the electronegativity decreased. A high percentage of C-N+ means that the enzyme is better able to bind to the functionalizing substance7,31. This appeared to follow an electronegativity trend.
(a) Diagram of the fabricated metal-oxide EGFETs-based glucose sensor device. The inset image shows a top view of the attached PDMS wall and the sensing method. (b) Characteristics of the transfer curves (IDS-VGS, Ag/AgCl) of InGaO (blue circle), InO (black circle), and InZnO (red circle) EGFETs-based devices in 1× PBS solution. (c) Characteristics of the transfer curves (IDS-VGS) of InGaO (blue circle), InO (black circle), and InZnO (red circle) devices in air.
Figure 2a shows diagrams and photographs of the fabricated metal-oxide EGFET-based glucose sensor devices. The devices were operated in 1× PBS, and a commercially available leak-free Ag/AgCl reference electrode placed over the sensing area was used as the floating gate. A PDMS wall was used on top of the exposed metal oxide channel to define the sensing area and contain the electrolyte.
The electrical characteristics of the device measured in the 1× PBS solution differed from those measured in air, as shown in Fig. 2b and c. In particular, the Ion/off ratio obtained in the air state had a value of approximately 107 to 108 at a drain voltage of 5 V and gate voltage in the range of -30–30 V. The transfer curves of each device indicate that we obtained high field effect mobilities (µFET) of 16.81, 17.89, and 9.15 cm2/Vs, respectively, in air. The threshold voltage (Vth) was calculated to be -2.15, -8.18, and − 1.08 V, respectively. On the other hand, the value obtained under 1× PBS solution was lower, ranging from 104 to 105 at a drain voltage of 0.1 V and gate voltage in the range of -0.5–0.5 V. This is related to the ionic strength of the 1× PBS solution and the low applied voltage. A comprehensive study was conducted by Kim et al.1 to investigate and analyze the effects of solvents on metal-oxide TFTs. The results of the study indicate that the solvent predominantly affects other donor-like defects, such as hydrogen donors and metal interstitials, or the change in deep traps originating from metal vacancies rather than oxygen vacancies as shallow donors. However, this is not a low (or poor) value for driving the device, and the sensitivity characteristics can be evaluated according to the molarity of glucose. Furthermore, as shown in Table 1, the device operation characteristics in the air and solution states fabricated by the sputtering system are superior, even when compared to the characteristics presented in previous papers. The transfer curves of each device indicate that they have high on/off ratios (> 104) and we obtained high field effect mobilities (µFET) of 8.7, 15.1, and 8.5 cm2/Vs, respectively, in 1× PBS. The subthreshold gate swing (SS) and threshold voltage (Vth) were calculated to be 40, 30, and 30 mV/dec. and 0.01, 0.06, and 0.03 V, respectively. The reproducibility of each device characteristic was excellent, and the results are shown in Fig. S3 and Table S3. The SS is an important parameter for these devices because it provides a gating effect deficiency and biosensor sensitivity13,28,32. The biosensor sensitivity is extremely high in the SS region owing to greater current changes, which is known as the gating effect owing to proton variation33. Therefore, to achieve optimal sensing performance, the devices were controlled by an Ag/AgCl reference electrode to tune their operating point to the SS regime. The mechanism of enzyme-based biosensors involves recognizing a biochemical reaction by attaching a recognition element to the device. Glucose detection is a representative application of enzyme-based sensors and byproducts in which the oxidation or reduction of the targets affects the electrical properties of the sensors. The structure used in this study is easy to fabricate, and the channel layer, which is exposed to the outside environment, can react more sensitively with biomolecules. An illustration of the glucose detection mechanism is presented in S4. Real-time measurements were performed according to changes in glucose concentration to determine the effect of differences in surface functionalization and enzyme imm obilization on the glucose detection characteristics, depending on the metal-oxide material used in the device. The enzymatic reactions between glucose and GOx are as follows9,34:
GOx catalyzes the conversion of glucose to gluconic acid and hydrogen peroxide (H2O2), which in turn react to form proton ions and electrons35. As the glucose concentration reacts with the GOx, the concentration of protons in the electrolyte increases. An increase in the proton concentration can control the charge in the channel layer of a metal-oxide EGFET-based glucose sensor device and amplify the current. The solution at each glucose concentration (~ 0.25µL) was injected into the PDMS wall (~ 100µL) using a micropipette. The glucose concentration was varied from 0 to 4 mM, corresponding to the tear (0.1–0.4 mM), urine (0–0.8 mM), sweat (0.01–1.11 mM), and saliva (8 µM–1.77 mM) glucose levels of diabetic patients. To quantify the sensor’s response to glucose, the real-time measurement results according to the glucose concentration, as shown in S5, were converted into the sensor’s response voltage change and the calibrated response voltage according to the glucose concentration (Fig. 3a and b).
(a) Response voltage and (b) calibrated response voltage according to glucose concentration of the InGaO, InO, and InZnO EGFET-based devices in 1× PBS.
The ΔVon plot according to glucose concentration is consistent with the Hill-Langmuir equation36 for equilibrium ligand-receptor binding:
where c is the glucose concentration, Kd is the dissociation constant of gluconic acid-glucose interaction, n is the Hill coefficient describing the cooperativity of the binding, ΔVon, max is the Von shift when gluconic acid molecules are bound to glucose37, and ΔVon, pristine is the Von in 1× PBS. As the electronegativity value decreased, the dissociation constants for the InGaO, InO, and InZnO EGFET-based glucose sensor devices increased to 0.15, 0.62, and 2.49 mM, respectively, and the glucose concentration corresponding to the highest calibrated response voltage value also increased.
By evaluating the transfer curve characteristics according to the glucose concentration, the calibrated response voltage was obtained using the following formula:
To assess its stability, the InO-based glucose sensor was subjected to additional testing using glucose titration at different intervals, as shown in S6. Although the response decreased slightly after the initial period, the relative sensitivity of the glucose sensor remained at a minimum of 94% during the first 10 d, similar to a previously published paper38. These results demonstrate that the glucose sensor can maintain a stable state over an extended period.
Table 1 shows the properties of the InGaO, InO, and InZnO EGFET-based glucose sensor devices compared with those of previously reported metal-oxide EGFET-based devices. The minimum detection range of the three samples was lower than that of the others, even for 1× PBS, which is a high-ionic-strength electrolyte39. The glucose concentration in different body secretions varies, offering choices for detecting glucose concentrations using specific secretions. The sensing characteristics of the InGaO (0.15 mM), InO (0.62 mM), and InZnO (2.49 mM) EGFET-based glucose sensors are comparable to those of other sensors that can detect glucose at very low concentrations in tears, urine, and sweat. Considering the constant dissociation characteristics of InGaO, InO, and InZnO, InGaO has the potential for use in sensors measuring the glucose concentration in tears, whose glucose concentration is relatively lower than that of other bodily secretions. InO appears to have the potential for use in sensors using urine or sweat, and InZnO may be applied to sensors using saliva or ISF. Therefore, materials that can react more sensitively to the concentration of glucose (specific region) in a specific detection substance can be selected. However, if devices with different compositions are used in parallel (such as devices using the materials introduced in this paper, that is, InO, InGaO, and InZnO), the linearity of the components can be achieved in a much broader range of areas. To confirm the effect of the electronegativity change caused by doping InO with Ga or Zn, we predicted the formation energies of surface functionalization and enzyme immobilization.
Reaction energy diagram at 0 eV (Ref.; InO) for the immobilization of the glucose oxidase (GOx) enzyme in three different samples.
The reaction energy diagrams for functionalization and enzyme immobilization are shown in Fig. 4. Owing to the large size of GOx (Aspergillus niger), we consider H-NH2 as GOx-NH2. In the first functionalization process, In(Ga/Zn)O was surface-hydroxylated by UV, and the surface was partially terminated with OH. These OH groups reacted with the C2H5 group from APTES, and APTES was attached to the In(Ga/Zn)O surface. Through cross-linking, the APTES-GDx chains were synthesized and terminated via C = O bonding. In the functionalization process, Ga doping (increasing electronegativity) decreased the sensitive formation energy at each step, whereas Zn doping (decreasing electronegativity) increased the formation energy at each step. Similar to the functionalization process, the formation energy of enzyme immobilization can be controlled by doping with Ga or Zn. At the end of the GDx, which is terminated by a C = O bond, GOx is immobilized to form a C = N-GOx (C = N-H by DFT calculations) bond. With Ga doping, the formation energy of immobilization becomes more negative than that of InO, whereas with Zn doping, the energy of immobilization becomes more positive than that of InO. Therefore, the amount of functionalized surface or immobilized enzyme should be in the order InZnO < InO < InGaO, which corresponds to the surface roughness and the number of N-N + bonds in XPS. As a control amount of GOx by Zn or Ga doping, the frequency of the reaction of d-glucose to glucose acid, as well as the concentration of glucose to which the oxide sensor responds (that is, becomes more sensitive or insensitive by Ga or Zn doping), can be modulated.
In this study, we induced changes in the surface termination of metal-oxide thin films by controlling their electronegativity using a sputtering deposition system. XPS analysis confirmed that the surface functionalization states of the modified thin films were different. The adhesion energy of the surface-functionalized material and the enzyme immobilization binding energy of the surface-functionalized thin film were calculated by DFT, and their glucose detection characteristics were evaluated. A difference in the amount of surface functionalization can change the material’s binding energy with glucose, resulting in a difference in the amount of charge it holds, thus indicating that the chemical potential energy of the attached glucose can be controlled. As the electronegativity increased, the percentage of positively charged protonated amine nitrogen (C-N+) increased, indicating that the binding of the enzyme to the functionalizing substance improved. In addition, a reduction in the formation energy for surface functionalization and enzyme immobilization resulted in a decrease in the dissociation constant. The sensitivity region of the glucose sensor could be adjusted by electronegativity control, and InGaO, InO, and InZnO exhibited differences in the most sensitive response regions at 0.15, 0.62, and 2.49 mM, respectively.
Therefore, materials that can react more sensitively to the concentration of glucose (specific region) present in a specific detection substance may be selected. In addition, the performance of glucose monitoring devices using these materials is comparable to that of devices fabricated using not only the sputtering system but also the spin coating, solution, and printing systems. To achieve practical applications and scalability of oxide-semiconductor-based biosensors, precise process control is crucial, and this can be implemented using a vacuum system. This study represents a milestone in the design of biomaterials. Furthermore, if devices with different compositions are used in parallel (such as those using materials introduced in this paper, that is, InO, InGaO, and InZnO), the linearity of the components can be achieved in a much broader range of areas.
The thermally grown 100-nm-thick silicon dioxide (SiO2) layer on a heavily doped p-type silicon (p++-Si) substrate was cleaned with acetone, ethanol, and DI water in a sonicator. After cleaning the substrate, the first shadow metal mask was attached to define the active channel layer. Then, a 10-nm-thick InO layer was deposited using a DC sputtering system with a 3-inch In2O3 (99.995%, 4N5) target at a power of 30 W without substrate temperature. After deposition, interdigitated Ti 30 nm (bottom) and Au 200 nm (top) electrodes were deposited using a thermal evaporation system. The source and drain areas were defined using a second shadow metal mask during electrode deposition. The fabricated channel width (W) and length (L) of the device were 10,200/200 µm, respectively. Finally, the InO EGFETs were annealed in a quartz tube furnace at 250 ℃ in air for 1 h to modify the stoichiometry of the films.
The fabrication methods for the InZnO and InGaO EGFETs were almost the same as those for the InO EGFETs, except for the co-sputtering method used in the deposition process of the active channel layer. Two 10-nm-thick InZnO EGFETs were fabricated, one using the co-sputtering of a 3-inch ZnO target (99.999%, 5 N, RF) at a power of 20 W and another using the co-sputtering of a 3-inch In2O3 target (99.995%, 4N5, DC) at a power of 30 W. Two 10-nm-thick InGaO EGFETs were fabricated, one using the co-sputtering of a 3-inch Ga2O3 target (99.99%, 4 N, RF) at a power of 20 W and another using the co-sputtering of a 3-inch In2O3 target at a power of 30 W without substrate temperature. The conditions used for the fabrication of all thin films were the same. The reactive gas flow rate ratio was Ar: O2 = 19:0.5 (O2 = 2.6%), and the working pressure of the main chamber was 5 mTorr.
InO-, InZnO-, and InGaO-FETs were UV-treated in air for 10 min before surface functionalization. The samples were modified stepwise for glucose detection. First, (3-aminopropyl)triethoxysilane (APTES, Sigma-Aldrich) was self-assembled on the surface of each device using 4 wt% APTES in toluene (Sigma-Aldrich) for 20 min. The samples were then directly rinsed with toluene. Next, the silanized samples were immersed in a 1.5 wt% glutaraldehyde linker (GDx, Sigma-Aldrich) in phosphate-buffered saline (1× PBS with pH 7.4, Sigma-Aldrich) solution for 5 min. The samples were immediately rinsed with 1× PBS. Next, 3 mg of glucose oxidase (GOx) from Aspergillus niger (Sigma-Aldrich) was immobilized on the APTES-GDx-treated surfaces in 6 ml of 1× PBS solution for 24 h. The samples were rinsed with 1× PBS. Finally, to protect the Au electrode from the electrolyte environment containing 1-dodecaneethiol (DDSH), the sample was treated with 1 wt% 1-DDSH in ethanol (99.99%, 4 N) for 1 h. All steps for surface functionalization were carried out at room temperature. The samples were then rinsed with ethanol. D-glucose (Sigma-Aldrich) was dissolved in 1× PBS at different concentrations.
The electrical characteristics and glucose-sensing performance of all the metal-oxide-semiconductor EGFET-based glucose sensors were evaluated at room temperature using a Keithley SCS-4200 semiconductor parameter analyzer and an Ag/AgCl leak-free reference electrode (LF-1, DONGWOO Science Co., Ltd.) in an electrolyte gate environment. For the evaluation of transistor’s performance, the transconductance (gm) of the metal oxide EGTFTs were calculated from an ID−VAg/AgCl curve using the following equation:
where W is the channel width, L is the channel length, and CDL is the electrical double layer capacitance per unit area in 0.1 M ionic strength aqueous solution reported previously (25.52 µF cm− 2)16. A comprehensive study was conducted Park et al.40 to investigate electrical double layer capacitance per unit area of metal oxide EGTFTs under various aqueous dielectric conditions. A polydimethylsiloxane (PDMS) solution was attached to the top of the device to determine the contact area of the analyte solution during the detection test according to the glucose concentration. The transfer curve characteristics (IDS-VGS) were measured at a fixed drain-to-source voltage (VDS) of 0.1 V. A low VDS was used to minimize electrochemical reactions that could occur when the source and drain electrodes come into contact with the analyte solution12. The sensitivity characteristics of metal-oxide-semiconductor EGFET-based glucose sensors according to glucose concentration were evaluated in real time in the following order. 1st 1x PBS (0.25µL) into the PDMS → measurement → removal of 1x PBS (0.25µL) from PDMS. 2nd Glucose mole concentration (0.25µL) into the PDMS → glucose detection → removal of glucose mole concentration (0.25µL) from PDMS → 1x PBS (0.25µL) into the PDMS to rinse for a few seconds. From the second step, it was performed repeatedly according to the change in glucose molar concentration.
Atomic Force Microscopy (AFM, Park Systems, NX10) was performed in an N2-purging glove box to minimize surface contamination and confirm the thin film’s topography. The chemical bonding states were determined by X-ray photoelectron spectroscopy (XPS) using a monochromatic Al Kα source with a pass energy of 29.35 eV. The XPS spectra were measured after eliminating the surface contamination caused by the adsorption of OH, C, H2O, etc., using Ar + ions at 500 eV. Transmission electron microscopy (TEM) images were obtained using a Talos F200X (FEI) equipped with an EDS (Bruker). The thickness of InGaO, InO, and InZnO thin films was determined using high-angle annular dark field (HAADF)-scanning transmission electron microscopy (STEM) images.
DFT calculations were performed using the Vienna Ab initio Simulation Package (VASP) with PBEsol functionality41,42,43. First, the bulk unit cell of cubic In2O3 was geometrically optimized with 3 × 3 × 3 K points and a cutoff energy of 500 eV until the condition of 0.02 eV/Å was met. Using the optimized bulk structure, we generated a (111) surface model with the lowest surface energy44. To minimize slab-to-slab interactions, a 50-Å slab of vacuum was inserted between the In2O3 layers. The various molecules related to functionalization and immobilization processes were attached to the In2O3 (111) surfaces, and the geometry was optimized using 3 × 3 × 1 K points and a cutoff energy of 500 eV until the 0.1 eV/Å condition was met. For the optimized structure, the total energies with 3 × 3 × 1 K points and a cutoff energy of 500 eV were predicted to calculate the formation energy of the reaction. To evaluate the effect of Ga/Zn doping, we randomly replaced 10% of In in the Ga or Zn system, optimized the geometry, and calculated the total energies under the same conditions as for In2O3.
The datasets used and/or analysed during the current study available from the corresponding author on reasonable request. Correspondence and requests for materials should be addressed to K.-B. Chung, K.-S. Jeong, or Y. S. Rim (email: [email protected]; [email protected]; [email protected]).
Kim, H. & Kwon, J. Y. Enzyme immobilization on metal oxide semiconductors exploiting amine functionalized layer. RSC Adv. 7, 19656–19661 (2017).
Article ADS CAS Google Scholar
Yoo, E. H. & Lee, S. Y. Glucose biosensors: an overview of use in clinical practice. Sens. (Basel). 10, 4558–4576 (2010).
Article ADS Google Scholar
So, C. F., Choi, K. S., Wong, T. K. S. & Chung, J. W. Y. Recent advances in noninvasive glucose monitoring. Med. Devices (Auckl). 5, 45–52 (2012).
CAS PubMed Google Scholar
Pambianco, G. et al. The 30-year natural history of type 1 diabetes complications: the Pittsburgh epidemiology of diabetes complications study experience. Diabetes. 55, 1463–1469 (2006).
Article CAS PubMed Google Scholar
Nathan, D. M. The diabetes control and complications trial/epidemiology of diabetes interventions and complications study at 30 years: overview. Diabetes Care. 37, 9–16 (2014).
Article CAS PubMed Google Scholar
Strakosas, X. et al. A non-enzymatic glucose sensor enabled by bioelectronic pH control. Sci. Rep. 9, 10844 (2019).
Article ADS PubMed PubMed Central Google Scholar
Toghill, K. E. & Compton, R. G. Electrochemical non-enzymatic glucose sensors: a perspective and an evaluation. Int. J. Electrochem. Sci. 5, 1246–1301 (2010).
Article CAS Google Scholar
Lee, H., Hong, Y. J., Baik, S., Hyeon, T. & Kim, D. H. Enzyme-based glucose sensor: from invasive to wearable device. Adv. Healthc. Mater. 7, e1701150 (2018).
Article PubMed Google Scholar
Bruen, D., Delaney, C., Florea, L. & Diamond, D. Glucose sensing for diabetes monitoring: recent developments. Sensors. 17, 1866 (2017).
Article ADS PubMed PubMed Central Google Scholar
Claussen, J. C. et al. Nano structuring platinum nanoparticles on multilayered graphene petal nanosheets for electrochemical biosensing. Adv. Funct. Mater. 22, 3399–3405 (2012).
Article CAS Google Scholar
Makaram, P., Owens, D. & Aceros, J. Trends in nanomaterial-based non-invasive diabetes sensing technologies. Diagnostics (Basel). 4, 27–46 (2014).
Article PubMed Google Scholar
Vu, C. A. & Chen, W. Y. Field-effect transistor biosensors for biomedical applications: recent advances and future prospects. Sens. (Basel). 19, 4214 (2019).
Article ADS CAS Google Scholar
Zhang, H. et al. Graphene-enabled wearable sensors for healthcare monitoring. Biosens. Bioelectron. 197, 113777 (2022).
Article CAS PubMed Google Scholar
Liu, N., Chen, R. & Wan, Q. Recent advances in electric-double-layer transistors for bio-chemical sensing applications. Sens. (Basel). 19, 3425 (2019).
Article ADS CAS Google Scholar
Rim, Y. S. Review of metal oxide semiconductors-based thin-film transistors for point-of-care sensor applications. J. Inf. Disp. 21, 203–210 (2020).
Article CAS Google Scholar
Rim, Y. S. et al. Printable ultrathin metal oxide semiconductor-based conformal biosensors. ACS Nano. 9, 12174–12181 (2015).
Article CAS PubMed Google Scholar
Fathollahzadeh, M. et al. Immobilization of glucose oxidase on ZnO nanorods decorated electrolyte-gated field effect transistor for glucose detection. J. Solid State Electrochem. 22, 61–67 (2018).
Bhat, K. S., Ahmad, R., Yoo, J. Y. & Hahn, Y. B. Nozzle-jet printed flexible field-effect transistor biosensor for high performance glucose detection. J. Colloid Interface Sci. 506, 188–196 (2017).
Article ADS CAS PubMed Google Scholar
Yano, M. et al. Zinc oxide ion-sensitive field-effect transistors and biosensors. Phys. Status Solidi A. 211, 2098–2104 (2014).
Article ADS CAS Google Scholar
Du, X., Li, Y., Motley, J. R., Stickle, W. F. & Herman, G. S. Glucose sensing using functionalized amorphous in – Ga – Zn – O field-effect transistors. ACS Appl. Mater. Interfaces. 8, 7631–7637 (2016).
Article CAS PubMed Google Scholar
Kim, H., Rim, Y. S. & Kwon, J. Y. Evaluation of metal oxide thin-film electrolyte-gated field effect transistors for glucose monitoring in small volume of body analytes. IEEE Sens. J. 20, 9004–9010 (2020).
CAS Google Scholar
Kim, Y. G. et al. Facile fabrication of wire-type indium gallium zinc oxide thin-film transistors applicable to ultrasensitive flexible sensors. Sci. Rep. 8, 5546 (2018).
Article ADS PubMed PubMed Central Google Scholar
Mohamad Nor, N., Ridhuan, N. S. & Abdul Razak, K. Progress of enzymatic and non-enzymatic electrochemical glucose biosensor based on nanomaterial-modified electrode. Biosensors. 12, 1136 (2022).
Article CAS PubMed PubMed Central Google Scholar
Naikoo, G. A. et al. Recent advances in non-enzymatic glucose sensors based on metal and metal oxide nanostructures for diabetes management-a review. Front. Chem. 9, 748957 (2021).
Article CAS PubMed PubMed Central Google Scholar
Bollella, P., Gorton, L., Ludwig, R. & Antiochia, R. A third generation glucose biosensor based on cellobiose dehydrogenase immobilized on a glassy carbon electrode decorated with electrodeposited gold nanoparticles: characterization and application in human saliva. Sens. (Basel). 17, 1912 (2017).
Article ADS Google Scholar
Teymourian, H., Barfidokht, A. & Wang, J. Electrochemical glucose sensors in diabetes management: an updated review (2010–2020). Chem. Soc. Rev. 49, 7671–7709 (2020).
Article CAS PubMed Google Scholar
Lipińska, W. et al. The optimization of enzyme immobilization at Au-Ti nanotextured platform and its impact onto the response towards glucose in neutral media. Mater. Res. Express. 6, 1150e3 (2019).
Article Google Scholar
Sarkar, D. & Banerjee, K. Proposal for tunnel-field-effect-transistor as ultra-sensitive and label-free biosensors. Appl. Phys. Lett. 100, 143108 (2012).
Article ADS Google Scholar
Desimoni, E. & Brunetti, B. X-ray photoelectron spectroscopic characterization of chemically modified electrodes used as chemical sensors and biosensors: a review. Chemosensors. 3, 70–117 (2015).
Article CAS Google Scholar
Major, G. H. et al. Practical guide for curve fitting in x-ray photoelectron spectroscopy. J. Vac Sci. Technol. A. 38, 061203 (2020).
Article CAS Google Scholar
BaO, S. J., Wang, Y., Wei, Z., Yang, W. & Yu, Y. Highly efficient recovery of heavy rare earth elements by using an amino-functionalized magnetic graphene oxide with acid and base resistance. Hazard. Mater. 424, 127370 (2022).
Article CAS Google Scholar
Li, B. R., Chen, C. C., Kumar, U. R. & Chen Y.-T. Advances in nanowire transistors for biological analysis and cellular investigation. Analyst. 139, 1589–1608 (2014).
Article ADS CAS PubMed Google Scholar
Gao, X. P. A., Zheng, G. & Lieber, C. M. Subthreshold regime has the optimal sensitivity for nanowire FET biosensors. Nano Lett. 10, 547–552 (2010).
Article ADS CAS PubMed PubMed Central Google Scholar
Yoon, H., Ko, S. & Jang, J. Field-effect-transistor sensor based on enzyme-functionalized polypyrrole nanotubes for glucose detection. J. Phys. Chem. B. 112, 9992–9997 (2008).
Article CAS PubMed Google Scholar
Kang, B. C., Park, B. S., Ha, T. J. & Ha & T.-J. highly sensitive wearable glucose sensor systems based on functionalized single-wall carbon nanotubes with glucose oxidase-nafion composites. Appl. Surf. Sci. 470, 13–18 (2019).
Article ADS CAS Google Scholar
Hill, A. V. A new mathematical treatment of changes of ionic concentration in muscle and nerve under the action of electric currents, with a theory as to their mode of excitation. J. Physiol. 40, 190–224 (1910).
Article ADS CAS PubMed PubMed Central Google Scholar
Chen, H. et al. Quasi-two-dimensional metal oxide semiconductors based ultrasensitive potentiometric biosensors. ACS Nano. 11, 4710–4718 (2017).
Article CAS PubMed Google Scholar
Huang, L., Jia, Z., Liu, H., Pi, X. & Zhou, J. Design of a sandwich hierarchically porous membrane with oxygen supplement function for implantable glucose sensor. Appl. Sci. 10, 2848 (2020).
Article CAS Google Scholar
Palazzo, G. et al. Detection beyond Debye’s length with an electrolyte-gated organic field-effect transistor. Adv. Mater. 27, 911–916 (2015).
Article CAS PubMed Google Scholar
Park, S. Sub-0.5 V highly stable aqueous salt gated Metal Oxide Electronics. Sci. Rep. 5, 13088 (2015).
Article ADS CAS PubMed PubMed Central Google Scholar
Kresse, G. & Furthmüller, J. Efficiency of ab-initio total energy calculations for metals and semiconductors using a plane-wave basis set. Comput. Mater. Sci. 6, 15–50 (1996).
Article CAS Google Scholar
Kresse, G. & Joubert, D. From ultrasoft pseudopotentials to the projector augmented-wave method. Phys. Rev. B. 59, 1758–1775 (1999).
Article ADS CAS Google Scholar
Perdew, J. P. et al. Restoring the density-gradient expansion for exchange in solids and surfaces. Phys. Rev. Lett. 100, 136406 (2008).
Article ADS PubMed Google Scholar
Walsh, A. & Catlow, C. R. A. Structure, stability and work functions of the low index surfaces of pure indium oxide and Sn-doped indium oxide (ITO) from density functional theory. J. Mater. Chem. 20, 10438–10444 (2010).
Article CAS Google Scholar
Download references
This work was supported by the National Research Foundation of Korea(NRF) grant funded by the Korea government(MSIT)(RS-2024-00352476).
Division of Physics and Semiconductor Science, Dongguk University, Seoul, 100-715, Republic of Korea
Aeran Song, Min Jung Kim, Dong-Joon Yi & Kwun-Bum Chung
Department of Physics, Ewha Womans University, Seoul, 03760, Republic of Korea
Soyeong Kwon & Dong-Wook Kim
Advanced Analysis and Data Center, Korea Institute of Science and Technology (KIST), Seoul, 02792, Republic of Korea
Seunghwan Kim, Jee-Hwan Bae & Soohyung Park
Department of Intelligent Mechatronics Engineering and Convergence Engineering for Intelligent Drone, Department of Semiconductor Systems Engineering, Sejong University, Seoul, 05006, Republic of Korea
You Seung Rim
Division of AI Semiconductor, Yonsei University, Wonju, 26493, Republic of Korea
Kwang-Sik Jeong
You can also search for this author in PubMed Google Scholar
You can also search for this author in PubMed Google Scholar
You can also search for this author in PubMed Google Scholar
You can also search for this author in PubMed Google Scholar
You can also search for this author in PubMed Google Scholar
You can also search for this author in PubMed Google Scholar
You can also search for this author in PubMed Google Scholar
You can also search for this author in PubMed Google Scholar
You can also search for this author in PubMed Google Scholar
You can also search for this author in PubMed Google Scholar
You can also search for this author in PubMed Google Scholar
K.-B. C., K.-S. J., and Y. S. R. designed the experiments. A. S. fabricated metal oxide semiconductor thin films and EGFET-based glucose sensors and analyzed the surface properties of the thin films and the electrical performance of EGFET-based devices. M. J. K. and D.-J. Y measured and analyzed the XPS spectra. S. K. and D. W. K. measured and analyzed the AFM. S. Kim, J. H. Bae, and S. Park measured and analyzed the TEM data. K.-S. J. and A. S. wrote the manuscript. All authors discussed the experimental results and commented on the manuscript. A. S. and M. J. K. wrote the revised manuscript as co-first author. All the authors have approved the final version of the manuscript.
Correspondence to You Seung Rim, Kwang-Sik Jeong or Kwun-Bum Chung.
The authors declare no competing interests.
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Below is the link to the electronic supplementary material.
Open Access This article is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License, which permits any non-commercial use, sharing, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if you modified the licensed material. You do not have permission under this licence to share adapted material derived from this article or parts of it. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by-nc-nd/4.0/.
Reprints and permissions
Song, A., Kim, M.J., Yi, DJ. et al. Control of sensitivity in metal oxide electrolyte gated field-effect transistor-based glucose sensor by electronegativity modulation. Sci Rep 14, 27251 (2024). https://doi.org/10.1038/s41598-024-76885-x
Download citation
Received: 07 July 2024
Accepted: 17 October 2024
Published: 08 November 2024
DOI: https://doi.org/10.1038/s41598-024-76885-x
Anyone you share the following link with will be able to read this content:
Sorry, a shareable link is not currently available for this article.
Provided by the Springer Nature SharedIt content-sharing initiative